Improved delivery of the Neuroform 3 stent: technical note
Andrew Stephen Ferrell1, Kiarash Golshani, Ali Zomorodi
1Department of Radiology, Duke University, Durham, North Carolina, USA.
Journal of Neurointerventional Surgery
|October 13, 2011
Summary
A simple modification to the Neuroform 3 stent delivery system improves its flexibility. This technique enhances navigation and increases successful stent deployment rates for wide-neck aneurysms.
Area of Science:
- Neurosurgery
- Interventional Neuroradiology
- Medical Device Engineering
Background:
- Intracranial stents are crucial for treating wide-neck aneurysms not amenable to coiling alone.
- Technical challenges with stent delivery systems persist, impacting treatment success.
- The Neuroform 3 stent system, widely used, has reported deployment failures despite modifications.
Purpose of the Study:
- To describe and evaluate a simple back-table modification of the Neuroform 3 stent delivery system.
- To assess the impact of this modification on stent deliverability and deployment success.
Main Methods:
- Retrospective review of a cerebrovascular database for all Neuroform 3 stent uses.
- Identification of cases where the described back-table modification was applied.
- Analysis of stent placement failures related to delivery system trackability.
Main Results:
- 139 Neuroform 3 stents were placed using the modified technique.
- A technical failure rate of 2.2% was observed, with only three instances of placement failure.
- The modification was associated with improved trackability and navigation.
Conclusions:
- A simple back-table modification can enhance the Neuroform 3 delivery system's flexibility.
- This technique may improve navigation through complex intracranial vasculature.
- The modification shows potential for increasing successful stent deployment rates in aneurysm embolization.