Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Biological Effects of Radiation02:59

Biological Effects of Radiation

All radioactive nuclides emit high-energy particles or electromagnetic waves. When this radiation encounters living cells, it can cause heating, break chemical bonds, or ionize molecules. The most serious biological damage results when these radioactive emissions fragment or ionize molecules. For example, α and β particles emitted from nuclear decay reactions possess much higher energies than ordinary chemical bond energies. When these particles strike and penetrate matter, they produce ions...
Scanning Electron Microscopy01:07

Scanning Electron Microscopy

A scanning electron microscope (SEM) is used to study the surface features of a sample by using an electron beam that scans the sample surface in a two-dimensional manner. Typically, areas between ~1 centimeter to 5 micrometers in width can be imaged. SEM can be used to image bacteria, viruses, tissues as well as larger samples like insects. Conventional SEM gives a magnification ranging from 20X to 30,000X and spatial resolution of 50 to 100 nanometers.
Fundamental Principles
Accelerated...

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Should reference beam data be used for beam modeling during linear accelerator commissioning?

Medical physics·2026
Same author

Monte Carlo modeling and phantom studies show Cherenkov emission per unit dose during total skin electron therapy is a function of tissue optical properties.

Medical physics·2026
Same author

3D Reactive Oxygen Species Dosimetry in Pleural Photodynamic Therapy: Integration of Macroscopic Kinetic Modeling and Deformable Registration.

Antioxidants (Basel, Switzerland)·2026
Same author

Evaluation of a time-resolved singlet oxygen detection system for in vivo photodynamic therapy.

Biomedical optics express·2026
Same author

Stathmin 1 regulates epileptic seizures by influencing GABA<sub>A</sub> R-mediated inhibitory synaptic activity.

Neurobiology of disease·2026
Same author

In Vivo Assessment of Benzoporphyrin Uptake and Singlet Oxygen Generation in Mice for Photodynamic Therapy Monitoring.

ACS photonics·2026
Same journal

Decomposition-based harmonization for quantitative PET imaging across scanners and radiotracers.

Medical physics·2026
Same journal

Development and evaluation of an in vivo dose-based monitoring system for electron FLASH radiation therapy.

Medical physics·2026
Same journal

A novel optical respiratory gating system with a hybrid phase-amplitude algorithm for spot-scanning proton therapy.

Medical physics·2026
Same journal

Gamma Knife treatment planning using knowledge-based reinforcement learning.

Medical physics·2026
Same journal

Development and characterization of a novel, small animal external beam irradiator using a clinical high dose rate brachytherapy source.

Medical physics·2026
Same journal

Deep learning-based dose prediction for MR-guided prostate SIB: Supporting rapid feasibility assessment and adaptive editing margin selection.

Medical physics·2026
See all related articles

Related Experiment Video

Updated: May 28, 2026

Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
07:31

Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator

Published on: May 9, 2014

Backscatter correction factor for megavoltage photon beam.

Yida Hu1, Timothy C Zhu

  • 1Department of Radiation Oncology, University of Pennsylvania, Philadelphia, PA, USA.

Medical Physics
|October 14, 2011
PubMed
Summary
This summary is machine-generated.

Determining adequate backscatter phantom thickness is crucial for accurate clinical dosimetry. This study establishes minimum backscatter thicknesses for 6 and 15 MV photon beams, ensuring full backscatter conditions for precise dose measurements.

More Related Videos

Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
08:34

Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies

Published on: February 6, 2019

Irradiator Commissioning and Dosimetry for Assessment of LQ &alpha; and &beta; Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
06:20

Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition

Published on: March 11, 2021

Related Experiment Videos

Last Updated: May 28, 2026

Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
07:31

Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator

Published on: May 9, 2014

Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
08:34

Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies

Published on: February 6, 2019

Irradiator Commissioning and Dosimetry for Assessment of LQ &alpha; and &beta; Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
06:20

Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition

Published on: March 11, 2021

Area of Science:

  • Medical Physics
  • Radiation Oncology
  • Radiotherapy Physics

Background:

  • Accurate clinical dosimetry for photon beams requires ensuring full backscatter conditions.
  • Insufficient backscatter thickness necessitates the use of backscatter correction factors (BCF).

Purpose of the Study:

  • To determine the minimum backscatter phantom material thickness for routine clinical dosimetry of photon beams.
  • To establish backscatter correction factors (BCF) for varying phantom thicknesses, field sizes, and photon beam energies.

Main Methods:

  • Experimental measurements were performed using a thimble chamber in SAD geometry for 6 and 15 MV photon beams.
  • A convolution method was employed to calculate BCF using Monte-Carlo simulated point-spread kernels.
  • Measurements covered field sizes from 10x10 cm to 30x30 cm and depths from d(max) to 20 cm.

Main Results:

  • Convolution calculations showed excellent agreement (within 0.8%) with experimental measurements.
  • The backscatter correction factor (BCF) deviates more from 1 with lower energies and larger field sizes.
  • Minimum BCF values were predicted at d(max) and 40x40 cm field size (0.970 for 6 MV, 0.983 for 15 MV).

Conclusions:

  • Recommended backscatter thicknesses are 6.0 cm for 6 MV and 4.0 cm for 15 MV for field sizes up to 10x10 cm (BCF=0.998).
  • Using 4 cm backscatter thickness results in BCF values of 0.997 (6 MV) and 0.998 (15 MV) for 10x10 cm fields.
  • For larger field sizes (40x40 cm), 4 cm backscatter thickness yields BCFs of 0.983 (6 MV) and 0.990 (15 MV).