Related Experiment Video
Updated: May 28, 2026

Multi-Modal Home Sleep Monitoring in Older Adults
Published on: January 26, 2019
Validation of a polyvinylidene fluoride impedance sensor for respiratory event classification during polysomnography
Brian B Koo1, Colin Drummond, Susan Surovec
1Case Western Reserve University School of Medicine, Department of Neurology, Cleveland, OH 44120, USA. koobri@gmail.com
Study Objectives:
The AASM has recommended specific sensors in measuring apnea and hypopnea based on published reliability and validity data. As new technology emerges, these guidelines will need revision. Polyvinylidene fluoride (PVDF) measures impedance and can be incorporated into a belt to approximate airflow and respiratory effort. We compared respiratory event detection using PVDF impedance belts (PVDFb), respiratory inductance plethysmography (RIP), and nasal-oral pneumotachography (PNT).
Methods:
First, in a clinical setting, 50 subjects (median AHI 26) undergoing polysomnography were fitted with PVDFb and standard sensors. Studies were scored in 4 independent passes using 4 respiratory montages (M); M1: nasal pressure transduction (NPT), thermistry, and RIP; M2: NPT, thermistry, and PVDFb; M3: thermistry and PVDFb; M4: PVDFb alone. Each experimental montage (M2-M4) was compared to the reference standard (M1) for total apneas and hypopneas. In a second experimental study, respiratory event detection was compared across a series of breathing trials for PVDFb, RIP, and PNT in normal subjects. Agreement was evaluated with intraclass correlation coefficient (ICC), κ statistics, and Bland-Altman plots.
Results:
ICCs comparing event numbers by M1 to M 2, 3, and 4 were: 0.99, 0.93, and 0.91, respectively. Almost identical numbers of events were identified for M 1 and M2 (177.5 ± 122.7 vs 177.6 ± 123.2). Event subtypes also were comparable. PVDFb was less sensitive than PNT but no different than RIP in detecting decreased breathing amplitude.
Conclusions:
PVDFb was comparable to standard RIP in determining respiratory events during polysomnography and in detecting decreased breathing amplitude, suggesting that PVDFb can be used as an alternative to RIP for apnea/hypopnea evaluation.
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement: