Pilot In Vitro Study to Assess Cleaning Ability and Effects of Different Decontamination Methods on Implant Surfaces
Published on: November 21, 2025
Time-dependent recontamination rates of sterilised dental instruments
1School of Oral and Dental Sciences, University of Bristol, Lower Maudlin Street, Bristol, BS1 2LY. chris.barker@bristol.ac.uk
Objectives:
To determine time-related recontamination rates of sterilised instruments, following guidance from the UK Department of Health (HTM 01-05) that such instruments within primary dental care may only be stored for 60 days following sterilisation using a vacuum autoclave.
Materials And Methods:
A total of 25 used examination mirrors underwent a washer-disinfector cycle, individual packaging and finally vacuum autoclaving. Immediately after autoclaving, time zero, five mirrors were tested for microbial contamination by aerobic and anaerobic culture. At 31, 60, 90 and 124 days a further five mirrors were removed from their packaging and were similarly tested for microbial contamination.
Results:
There was no bacterial growth on blood-enriched media under both aerobic and anaerobic conditions after 5 days of incubation at 37°C at any time period from 0 to 124 days post-sterilisation.
Conclusions:
There was no recontamination of sterilised instruments in this investigation over the test period of 124 days. This exceeds the recommended limit of 60 days stated by the UK Department of Health. The new guidance, HTM 01-05, appears to place an extra burden on primary care dentists. This burden is not without associated costs, and at present does not appear to be based on published evidence.
Related Concept Videos
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Hand hygiene
Hand washing...