Impact of heart rate and rhythm on radiation exposure in prospectively ECG triggered computed tomography
Christian Luecke1, Claudia Andres, Borek Foldyna
1University of Leipzig-Heart Center, Department of Diagnostic and Interventional Radiology, Strümpellstrasse 39, D-04289, Leipzig, Germany. neep@gmx.de
Purpose:
To evaluate the influence of different heart rates and arrhythmias on scanner performance, image acquisition and applied radiation exposure in prospectively ECG triggered computed tomography (pCT).
Materials And Methods:
An ECG simulator (EKG Phantom 320, Müller & Sebastiani Elektronik GmbH, Munich, Germany) was used to generate different heart rhythms and arrhythmias: sinus rhythm (SR) at 45, 60, 75, 90 and 120/min, supraventricular arrhythmias (e.g. sinus arrhythmia, atrial fibrillation) and ventricular arrhythmias (e.g. ventricular extrasystoles), pacemaker-ECGs, ST-changes and technical artifacts. The analysis of the image acquisition process was performed on a 64-row multidetector CT (Brilliance, Philips Medical Systems, Cleveland, USA). A prospectively triggered scan protocol as used for routine was applied (120 kV; 150 mAs; 0.4s rotation and exposure time per scan; image acquisition predominantly in end-diastole at 75% R-R-interval, in arrythmias with a mean heart rate above 80/min in systole at 45% of the R-R-interval; FOV 25 cm). The mean dose length product (DLP) and its percentage increase from baseline (SR at 60/min) were determined.
Result:
Radiation exposure can increase significantly when the heart rhythm deviates from sinus rhythm. ECG-changes leading to a significant DLP increase (p<0.05) were bifocal pacemaker (61%), pacemaker dysfunction (22%), SVES (20%), ventricular salvo (20%), and atrial fibrillation (14%). Significantly (p<0.05) prolonged scan time (>8 s) could be observed in bifocal pacemaker (12.8 s), pacemaker dysfunction (10.7 s), atrial fibrillation (10.3 s) and sinus arrhythmia (9.3 s).
Conclusion:
In prospectively ECG triggered CT, heart rate and rhythm can provoke different types of scanner performance, which can significantly alter radiation exposure and scan time. These results might have an important implication for indication, informed consent and contrast agent injection protocols.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage. When...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...