Retinal oximeter for the blue-green oximetry technique
Kurt R Denninghoff1, Katarzyna B Sieluzycka, Jennifer K Hendryx
1University of Arizona, Department of Emergency Medicine and College of Optical Sciences, 1609 North Warren Avenue, Room 116, Tucson, Arizona 85724-5057, USA. kdenninghoff@aemrc.arizona.edu
Journal of Biomedical Optics
|October 28, 2011
Summary
This study introduces a novel retinal oximetry technique, blue-green oximetry (BGO), achieving accurate, noninvasive measurement of blood oxygen saturation. The developed device demonstrates clinical utility by meeting the critical ±3% accuracy requirement for retinal oximetry.
Area of Science:
- Ophthalmology
- Biomedical Engineering
- Medical Imaging
Background:
- Noninvasive assessment of central venous oxyhemoglobin saturation (SO(2)) using retinal oximetry is clinically valuable.
- Previous retinal oximetry designs lacked the required ±3% saturation calibration accuracy for clinical use.
Purpose of the Study:
- To demonstrate the feasibility of the blue-green oximetry (BGO) technique for accurate, noninvasive retinal oximetry.
- To develop and test a noninvasive on-axis scanning retinal oximeter (ROx-3) capable of achieving clinical accuracy.
Main Methods:
- Development of a noninvasive on-axis scanning retinal oximeter (ROx-3) utilizing the blue-green oximetry (BGO) technique.
- Implementation of a field stop and anticonfocal bisecting wire to isolate multiply backscattered light and remove specular reflections.
- Testing the device on enucleated swine eye vessels and a human retinal vein.
Main Results:
- The blue-green oximetry (BGO) technique achieved accuracy within ±3% in swine.
- Retinal SO(2) measurements in human volunteers were internally consistent with a standard error of the mean of ±2% SO(2).
- Absolute saturation measures were within the expected physiological ranges.
Conclusions:
- The developed noninvasive on-axis scanning retinal oximeter (ROx-3) successfully implements the blue-green oximetry (BGO) technique.
- This study represents the first demonstration of noninvasive on-axis BGO retinal oximetry.
- The achieved accuracy meets the clinical requirement for assessing retinal oxyhemoglobin saturation.
Related Concept Videos
Pulse Oximetry
Pulse oximetry, or SpO2, is a non-invasive method for continuously monitoring arterial oxygen saturation (SaO2). This procedure involves attaching a probe or sensor to the patient's fingertip, forehead, earlobe, or nose bridge. The sensor works by detecting changes in oxygen saturation levels through light signals generated by the oximeter and reflected by the pulsing blood under the probe.
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Following these guidelines can help nurses accurately measure vital signs, assess changes in patient conditions, and provide timely treatment when necessary. Adhering closely to the guidelines ensures the accuracy and reliability of the results.
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.