Urinary metanephrines by liquid chromatography tandem mass spectrometry: using multiple quantification methods to
Zlatuse D Clark1, Elizabeth L Frank
1ARUP Institute for Clinical and Experimental Pathology, Salt Lake City, UT 84108, USA.
A new liquid chromatography-tandem mass spectrometry (LC-MS/MS) method accurately measures urinary metanephrines, improving pheochromocytoma diagnosis. This high-throughput assay overcomes common interferences, offering reliable results for clinical use.
Area of Science:
- Clinical Chemistry
- Analytical Chemistry
- Mass Spectrometry
Background:
- Accurate determination of urinary metanephrines is crucial for diagnosing and monitoring pheochromocytoma.
- Existing assays for metanephrines are frequently hindered by interferences, impacting reliability.
- A need exists for a robust, high-throughput method for urinary metanephrine and normetanephrine measurement.
Purpose of the Study:
- To develop and validate a reliable, high-throughput liquid chromatography-tandem mass spectrometry (LC-MS/MS) method.
- To measure metanephrine and normetanephrine concentrations in urine samples.
- To address and overcome interferences common in existing assays.
Main Methods:
- Urine samples underwent acid hydrolysis followed by solid-phase extraction using a mixed-mode cation exchange sorbent in a 96-well format.
- Separation was achieved using a Restek perfluorophenyl column with an isocratic mobile phase and gradient cleanout.
- Detection was performed on an API 3200 triple quadrupole mass spectrometer using electrospray ionization in positive mode and multiple reaction monitoring.
Main Results:
- The method demonstrated linearity up to 15,000 nmol/L with limits of detection and quantification at 2.5 and 10 nmol/L, respectively.
- Imprecision values (within-run, between-day, total) were consistently low, at or below 2.7% for both analytes.
- The LC-MS/MS method showed excellent correlation with the established GC-MS method, featuring a 6-minute injection-to-injection time.
Conclusions:
- The developed LC-MS/MS method is a reliable and high-throughput assay for quantifying urinary metanephrine and normetanephrine.
- This validated method effectively overcomes interferences, providing accurate results for pheochromocytoma diagnosis.
- The method has been successfully implemented in clinical service since August 2010, demonstrating exceptional performance.
More Related Videos
11:00Untargeted Metabolomics from Biological Sources Using Ultraperformance Liquid Chromatography-High Resolution Mass Spectrometry (UPLC-HRMS)
Published on: May 20, 2013
08:41Integration of Miniaturized Solid Phase Extraction and LC-MS/MS Detection of 3-Nitrotyrosine in Human Urine for Clinical Applications
Published on: July 14, 2017
Related Concept Videos
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
High-Performance Liquid Chromatography: Types of Detectors