Related Experiment Video
Updated: May 28, 2026

Impedance Pneumography for Minimally Invasive Measurement of Heart Rate in Late Stage Invertebrates
Published on: April 4, 2020
Effects of temperature and time delay on arterial blood gas and electrolyte measurements
Panida Srisan1, Teerasak Udomsri, Pravit Jetanachai
1Department of Pediatrics, Queen Sirikit National Institute of Child Health, College of Medicine, Rangsit University, Bangkok, Thailand. psrisan@yahoo.com
Objective:
To determine the changes in pH, PaO2, PaCO2 and Na, K, Cl in arterial blood samples stored at room temperature or on ice, at 0, 15, 30, 45 and 60 minutes.
Material And Method:
Arterial blood samples were collected in heparinized capillary tubes and stored at room temperature (24-26 degrees C) and on ice (0-4 degrees C). ABG and electrolytes were measured at 0, 15, 30, 45 and 60 minute intervals.
Results:
There were significant decreases in the pH, PaO2, Na, Cl and significant increases in PaCO2 and K over time in both groups. The changes were greater and faster at room temperature. The significant decrease in pH over time was not found until 30 minutes at room temperature and 45 minutes on ice. There were significant decreases in PaO2, concurrent with significant increases in PaCO2 from 15 minutes onwards in both groups. Both Na and K exhibited a significant change at 60 minutes in the room temperature group. Significant decreases of Cl over time were not found until 15 minutes at room temperature, and 30 minutes on ice.
Conclusion:
For ABG and electrolytes analysis, the blood sample should be analyzed within 15 minutes and be stored at either room temperature or on ice.
More Related Videos
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to reabsorb or...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Errors occurring during blood pressure monitoring
Several factors...
Disorders of Acid-Base Balance
Respiratory Acidosis and Alkalosis
Respiratory acidosis occurs due to an increase in the partial pressure of carbon dioxide PCO2 in the blood. It often arises from shallow breathing or impaired gas exchange caused by...