[Reliability of the pulse contour analysis for cardiac output measurement for assessing the fluid responsiveness]
V Reubrecht1, G Dingemans, A Gache
1Service des réanimations, division anesthésie réanimation douleur urgence, CHU de Nîmes, place du Pr. Robert-Debré, 30029 Nîmes cedex 9, France.
Background:
The cardiac output (CO) is classically measured in intensive care unit patients. pulse contour (PC) method allows monitoring of CO.
Objective:
The aim of the present study was to assess the ability of PC to assess the fluid responsiveness defined as an increase in CO more than or equal to 15% after 500 mL hydroxyethyl starch over 20 minutes.
Patients And Methods:
In this observational prospective study, patients in shock in whom a PC method was inserted were included. CO was measure using the PC and thermodilution methods before and after a fluid challenge indicated by the physician. The correlation coefficient was measured, the diagram of Bland and Altman was built and the percentage of error (Critchley and Critchley method) was calculated. The ability of PC to diagnose fluid responsiveness was assessed using a receiver operating characteristics (ROC) curve.
Results:
Sixty-two fluid challenges were performed in 37 included patients. After fluid challenge, r(2) was 0.05 (P<0.01), the bias between PC and thermodilution was 0.3 ± 1.2L/min and the percentage of error was 36%. The area of the ROC curve was 0.601 [0.468-0.723].
Conclusion:
In ICU patients with shock, PC cannot replace thermodilution to diagnose fluid responsiveness.
Related Concept Videos
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls between...
Pulse
The pulse serves as a clinical indicator...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart rate...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac output averages...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...