Related Experiment Video
Updated: May 27, 2026

03:47
Enhancement of Facial Rejuvenation Through a Combination of 1565 nm Non-Ablative Fractional Laser with 30% Supramolecular Salicylic Acid
Published on: September 27, 2024
Multiple facial burns with the new Thermage CPT system
Flor A Mayoral1, Janelle M Vega
1Flor A. Mayoral Dermatology Group, Miami, FL, USA.
Journal of Drugs in Dermatology : JDD
|November 5, 2011
Summary
A patient experienced facial burns from the Thermage CPT system due to a device malfunction. This suggests Thermage CPT may be contraindicated for male patients seeking skin tightening treatments.
Area of Science:
- Dermatology
- Medical Device Technology
Background:
- Thermage CPT system is a non-invasive radiofrequency device for skin tightening.
- Patient presented for cosmetic skin tightening procedures.
Observation:
- A 45-year-old male patient underwent Thermage CPT treatment.
- The patient sustained multiple facial burns during the procedure.
Findings:
- The burns were attributed to a failure of the device's dielectric membrane.
- This specific incident highlights a potential safety concern with the Thermage CPT system.
Implications:
- The Thermage CPT system may pose a contraindication for treating male patients.
- Further investigation into device safety and patient selection is warranted for radiofrequency skin tightening procedures.