A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
An appropriate defibrillation threshold obtained by the combined connection between two shock leads and ICD generator
Akinori Sato1, Masaomi Chinushi, Kenichi Iijima
1Division of Cardiology, Niigata University Graduate School of Medical and Dental Sciences, Japan.
Abstract:
A 60-year-old man with arrhythmogenic right ventricular cardiomyopathy was readmitted for the battery exchange of his implantable cardioverter-defibrillator (ICD). Since (i) he had been treated with a dual-coil shock lead (Sprint Fidelis, Medtronic) and (ii) pre-operative venography showed mild collateral flow to the left subclavian vein, a single-coil lead was additionally implanted. However, the single-coil defibrillation system was unable to terminate the induced ventricular fibrillation (VF), thus dual defibrillation shock pathways were created using the connection to the superior vena cava coil of the Fidelis lead. The combined connections of the two shock leads provided an appropriate margin of the defibrillation threshold.
Related Concept Videos
Cardiopulmonary Resuscitation III: AED Use
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Cardiopulmonary Resuscitation IV: Pharmacological Management
Insulation Coordination