Electromagnetic interference in a cardiac pacemaker during cauterization with the coagulating, not cutting mode
Basem Abdelmalak1, Narasimhan Jagannathan, Faisal D Arain
1Department of General Anesthesiology and Outcomes Research, Cleveland Clinic, Cleveland, OH, USA.
Electrosurgery in the coagulation mode can cause dangerous pacemaker interference, leading to asystole. Careful management and awareness are crucial for patient safety during procedures involving pacemakers.
Area of Science:
- Cardiology
- Biomedical Engineering
- Neurosurgery
Background:
- Monopolar electrocautery is commonly used in surgery, with electromagnetic interference (EMI) typically associated with its cutting mode.
- Dual-chamber, rate-modulated pacemakers (DDDR) are susceptible to EMI, potentially causing hemodynamic instability.
- Understanding EMI risks is vital for patient safety during procedures involving electronic implants.
Observation:
- A case of a 74-year-old male patient with a DDDR pacemaker experiencing intraoperative EMI during a spine procedure.
- The EMI occurred specifically during the coagulation mode of monopolar electrocautery, not the cutting mode.
- The electrodispersive pad placement varied, but EMI persisted during coagulation, leading to asystole.
Findings:
- Electrocautery in the coagulation mode can induce significant EMI in DDDR pacemakers.
- This interference can lead to life-threatening events such as asystole.
- The location of the electrodispersive pad did not prevent EMI during coagulation mode.
Implications:
- Clinicians must be aware that the coagulation mode poses a risk for pacemaker EMI and hemodynamic compromise.
- Heightened vigilance and preparedness are essential for managing patients with pacemakers during electrosurgery.
- Implementing strategies to mitigate EMI risk can ensure favorable patient outcomes in neurosurgical and other procedures.
More Related Videos
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Related Concept Videos
Cardiac Action Potential
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials
Mechanism of Cardiac Arrhythmias
Electrophysiology of Normal Cardiac Rhythm
Cardiac Catheterization IV: Nursing Management
Cardiac Catheterization I: Pre-Procedure Overview
Cardiac Catheterization II: Right Heart Catheterization