Related Experiment Video
Updated: May 27, 2026

Remote Magnetic Navigation for Accurate, Real-time Catheter Positioning and Ablation in Cardiac Electrophysiology Procedures
Published on: April 21, 2013
Electromagnetic interference in a cardiac pacemaker during cauterization with the coagulating, not cutting mode
Basem Abdelmalak1, Narasimhan Jagannathan, Faisal D Arain
1Department of General Anesthesiology and Outcomes Research, Cleveland Clinic, Cleveland, OH, USA.
Abstract:
Electromagnetic interference in pacemakers has almost always been reported in association with the cutting mode of monopolar electrocautery and rarely in association with the coagulation mode. We report a case of electrocautery-induced electromagnetic interference with a DDDR pacemaker (dual-chamber paced, dual-chamber sensed, dual response to sensing, and rate modulated) in the coagulating and not cutting mode during a spine procedure. We also discuss the factors affecting intraoperative electromagnetic interference. A 74-year-old man experienced intraoperative electromagnetic interference that resulted in asystole caused by surgical electrocautery in the coagulation mode while the electrodispersive pad was placed at different locations and distances from the operating site (This electromagnetic interference did not occur during the use of the cutting mode). However, because of careful management, the outcome was favorable. Clinicians should be aware that the coagulation mode of electrocautery can cause electromagnetic interference and hemodynamic instability. Heightened vigilance and preparedness can ensure a favorable outcome.
More Related Videos
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Related Concept Videos
Cardiac Action Potential
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials
Mechanism of Cardiac Arrhythmias
Electrophysiology of Normal Cardiac Rhythm
Cardiac Catheterization IV: Nursing Management
Cardiac Catheterization I: Pre-Procedure Overview
Cardiac Catheterization II: Right Heart Catheterization