Related Experiment Video
Updated: May 27, 2026

Determination of the Gas-phase Acidities of Oligopeptides
Published on: June 24, 2013
[Determination of four acids in the air of workplace by gas chromatography]
1Jiangsu Center for Disease Control and Prevention, Nanjing 210009, China.
Objective:
Four kinds of acids (acetic acid, propionic acid, acrylic acid and methylacrylic acid) in the air in a workplace were quantitatively determined by gas chromatography synchronously.
Methods:
Four acids in the air were adsorbed by silica gel tube sampling and solvent desorption using acetone, then analyzed by GC with FFAP capillary column.
Results:
To acetic acid, propionic acid, acrylic acid and methylacrylic acid, the linear regression equations were respectively y = -4.3+1.46x (r = 0.999), y = 0.4+2.37x (r = 0.999), y = 10.4+1.73x (r = 0.999) and y = -2.3+3.21x (r = 0.999). The detection limits were respectively 3.4 microg/mL, 2.1 microg/mL, 2.9 microg/mL and 1.6 microg/mL. The average desorption efficiencies were respectively 92.2%-92.8%, 94.1%-97.4%, 94.8%-95.4% and 94.1%-98.3%. The relative standard deviations were 1.1%-4.0%, 1.2%-7.8%, 0.9%-4.0% and 1.6%-4.8%.
Conclusion:
The method is suitable to determine four kinds of acids in the air in a workplace.
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...