Noncontrast MR angiography for supraaortic arteries using inflow enhanced inversion recovery fast spin echo imaging
Naoyuki Takei1, Mitsuharu Miyoshi, Hiroyuki Kabasawa
1Applied Science Laboratory, GE Healthcare, Tokyo, Japan. naoyuki.takei@ge.com
Purpose:
To depict supraaortic arteries using (3D fast spin echo (FSE) combined with slab selective inversion recovery for noncontrast magnetic resonance angiography (MRA) in healthy volunteers, and to investigate the property of the inflow enhanced inversion recovery FSE (IFIR-FSE) for background suppression and inflow effects.
Materials And Methods:
IFIR-FSE with no image subtraction was used to visualize the aortic arch, subclavian arteries, carotid arteries, and vertebral arteries in 10 healthy volunteers. Simulations were performed to achieve both high background suppression and inflow effects by adjusting the inversion time (TI) and wait time after data acquisition. The effect of inflow was investigated with TIs of 800, 1200, 1600, and 2000 msec. Contrast between artery and these background tissues were measured with scan protocols based on simulation results.
Results:
IFIR-FSE images showed good visualization of the supraaortic arteries and allowed separation of arteries from veins without image subtraction. The proposed method demonstrated that a high contrast between arteries and background tissues can be acquired with various TIs, which was in good agreement with the simulation. A TI over 1600 msec was favorable in terms of background suppression, arterial signal intensity, and inflow effects.
Conclusion:
Inflow enhanced inversion recovery cardiac-triggered 3D FSE imaging can be used for supraaortic artery imaging without contrast agents.
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...