Clinical application of Ultracontrol®: infusion volume and use with different dialyzers
M Albalate Ramón1, R Pérez García, P de Sequera Ortiz
1Nephrology, Hospital Infanta Leonor, Avenida Gran Vía del Este 80, Madrid, Spain. malbalater@senefro.org
Introduction:
Recent studies indicate that the survival benefit with post-dilution on line hemodiafiltration (OL-HDF-post) are achieved if the infusion volume (Vinf) is greater than 20L per session, a goal that is not easily achieved due to hemoconcentration problems. Today we have automated techniques to achieve higher performance minimizing the number of alarms as Ultracontrol® (Ultrac). The objective in the first part of study was to evaluate the UltraC performance (expressed as the filtration fraction (FF) and Vinf) and which problems it presents, and in the second part, to study its performance with four different dialyzers.
Material And Methods:
1st period. Nine patients were transferred to OL-HDF-post with UltraC. The first 3 months on OL-HDF all sessions were recorded and compared with hemodialysis sessions in the previous month. 2nd part: 18 patients on chronic OL-HDF-post were dialyzed for a week with each of these dialyser: FX1000, FX800, Elisio210H and Polyflux210.
Results:
1st period: In 3 patients, problems associated with inappropriate pressures emerged. In 3 patients there were problems associated with inadequate PTM and Psist that resolved changing to pressure control. Mean values were: maximum Qb 441 (21) (range 350-490) ml/min, Vinf 26.3 (3.3) l/session, FF 30.6 (2.5)%, KT 59.9 (5) l/session. KT increase of 12% compared to HD. 2nd part: Polyflux210 required less UltraC withdrawals than the others. Different PTM or Psist were found and determined the need for removal of the system. The KT was adequate. a) The UltraC system reaches FF of 30% with minimal alarms and Vinf higher than 20 l. b) Structural characteristics of dialysers can limit their use with UltraC although they managed to desirable KT and Vinf in a manual way.
Related Concept Videos
Extracorporeal Removal of Drugs: Continuous Renal Replacement Therapy
Continuous Renal Replacement Therapy
Peritoneal Dialysis I: Introduction and Procedure
Drug Dosing in Renal Diseases: Dose Adjustments Based on Drug Clearance and Elimination Rate Constant
Two-Compartment Open Model: IV Infusion
The model illustrates the decrease in plasma drug concentration from the central compartment with a specific equation. It shows that under steady-state conditions, the drug's input rate...
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...