Related Experiment Video
Updated: May 27, 2026

Drug-Induced Sleep Endoscopy (DISE) with Target Controlled Infusion (TCI) and Bispectral Analysis in Obstructive Sleep Apnea
Published on: December 6, 2016
The performance of two automatic servo-ventilation devices in the treatment of central sleep apnea
Shahrokh Javaheri1, Mark G Goetting, Rami Khayat
1University of Cincinnati College of Medicine, Medical Director Sleepcare Diagnostics, Cincinnati, OH 45040, USA. javaheri@snorenomore.com
Introduction:
This study was conducted to evaluate the therapeutic performance of a new auto Servo Ventilation device (Philips Respironics autoSV Advanced) for the treatment of complex central sleep apnea (CompSA). The features of autoSV Advanced include an automatic expiratory pressure (EPAP) adjustment, an advanced algorithm for distinguishing open versus obstructed airway apnea, a modified auto backup rate which is proportional to subject's baseline breathing rate, and a variable inspiratory support. Our primary aim was to compare the performance of the advanced servo-ventilator (BiPAP autoSV Advanced) with conventional servo-ventilator (BiPAP autoSV) in treating central sleep apnea (CSA).
Study Design:
A prospective, multicenter, randomized, controlled trial.
Setting:
Five sleep laboratories in the United States.
Participants:
Thirty-seven participants were included.
Measurements And Results:
All subjects had full night polysomnography (PSG) followed by a second night continuous positive airway pressure (CPAP) titration. All had a central apnea index ≥ 5 per hour of sleep on CPAP. Subjects were randomly assigned to 2 full-night PSGs while treated with either the previously marketed autoSV, or the new autoSV Advanced device. The 2 randomized sleep studies were blindly scored centrally. Across the 4 nights (PSG, CPAP, autoSV, and autoSV Advanced), the mean ± 1 SD apnea hypopnea indices were 53 ± 23, 35 ± 20, 10 ± 10, and 6 ± 6, respectively; indices for CSA were 16 ± 19, 19 ± 18, 3 ± 4, and 0.6 ± 1. AutoSV Advanced was more effective than other modes in correcting sleep related breathing disorders.
Conclusions:
BiPAP autoSV Advanced was more effective than conventional BiPAP autoSV in the treatment of sleep disordered breathing in patients with CSA.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Hyperpnea and Hyperventilation