Related Experiment Video
Updated: May 26, 2026

Automated Microfluidic Blood Lysis Protocol for Enrichment of Circulating Nucleated Cells
Published on: December 31, 2009
Microfluidic lysis of human blood for leukocyte analysis using single cell impedance cytometry
Xiaojun Han1, Cees van Berkel, James Gwyer
1School of Electronics and Computer Sciences, University of Southampton, Southampton, SO17 1BJ United Kingdom.
Abstract:
This paper demonstrates an integrated microfluidic system that performs a full blood count using impedance analysis. A microfluidic network design for red blood cell (RBC) lysis is presented, and the diffusive mixing processes are analyzed using experimental and simulated results. Healthy and clinical bloods analyzed with this system, and the data shows good correlation against data obtained from commercial hematology machines. The data from the microfluidic system was compared against hospital data for 18 clinical samples, giving R(2) (coefficient of determination) values of 0.99 for lymphocytes, 0.89 for monocytes, and 0.99 for granulocytes in terms of relative counts and 0.94 for lymphocytes, 0.91 for monocytes, and 0.95 for granulocytes in terms of absolute counts. This demonstrates the potential clinical utility of this new system for a point-of-care purpose.