Related Experiment Video
Updated: May 26, 2026

Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Intraoperative performance and postoperative outcome comparison of longitudinal, torsional, and transversal
Panos G Christakis1, Rosa M Braga-Mele
1Yale University School of Medicine, New Haven, Connecticut, USA.
Purpose:
To compare the intraoperative performance and postoperative outcomes of 3 phacoemulsification machines that use different modes.
Setting:
Kensington Eye Institute, Toronto, Ontario, Canada.
Design:
Comparative case series.
Methods:
This chart and video review comprised consecutive eligible patients who had phacoemulsification by the same surgeon using a Whitestar Signature Ellips-FX (transversal), Infiniti-Ozil-IP (torsional), or Stellaris (longitudinal) machine.
Results:
The review included 98 patients. Baseline characteristics in the groups were similar; the mean nuclear sclerosis grade was 2.0 ± 0.8. There were no significant intraoperative complications. The torsional machine averaged less phacoemulsification needle time (83 ± 33 seconds) than the transversal (99 ± 40 seconds; P=.21) or longitudinal (110 ± 45 seconds; P=.02) machines; the difference was accentuated in cases with high-grade nuclear sclerosis. The torsional machine had less chatter and better followability than the transversal or longitudinal machines (P<.001). The torsional and longitudinal machines had better anterior chamber stability than the transversal machine (P<.001). Postoperatively, the torsional machine yielded less central corneal edema than the transversal (P<.001) and longitudinal (P=.04) machines, corresponding to a smaller increase in mean corneal thickness (torsional 5%, transversal 10%, longitudinal 12%; P=.04). Also, the torsional machine had better 1-day postoperative visual acuities (P<.001).
Conclusions:
All 3 phacoemulsification machines were effective with no significant intraoperative complications. The torsional machine outperformed the transversal and longitudinal machines, with a lower mean needle time, less chatter, and improved followability. This corresponded to less corneal edema 1 day postoperatively and better visual acuity.