Related Experiment Video
Updated: Aug 15, 2026

Integrated Photoacoustic Ophthalmoscopy and Spectral-domain Optical Coherence Tomography
Published on: January 15, 2013
Dosimetry of Sr-90 ophthalmic applicators
C S Reft1, F T Kuchnir, I Rosenberg
1Department of Radiation and Cellular Oncology, University of Chicago, Illinois 60637.
Abstract:
Sr-90 ophthalmic applicators are commonly used for the treatment of superficial eye disorders. Although a variety of dosimetric devices such as film, thermoluminescent dosimeters (TLD's), ion chambers, and radiochromic foils have been used to measure the peak dose at the applicator surface, there is no internationally agreed upon calibration procedure. Recently, large discrepancies among calibrations of the same applicator at three institutions have been reported. Here we describe a technique to obtain the peak dose rate at the applicator surface using LiF TLD's. The technique can be used for the calibration of flat as well as curved surface applicators. Results for two flat and three concave applicators are presented. Our measurement of the surface dose rate for one of the flat applicators is compared with those obtained by four other institutions, each using different dosimetric devices.
More Related Videos
06:21Diffuse Optical Spectroscopy for the Quantitative Assessment of Acute Ionizing Radiation Induced Skin Toxicity Using a Mouse Model
Published on: May 27, 2016
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021