Related Experiment Video
Updated: May 26, 2026

All-electronic Nanosecond-resolved Scanning Tunneling Microscopy: Facilitating the Investigation of Single Dopant Charge Dynamics
Published on: January 19, 2018
Proton and iron ion observations from a solid-state microdosimeter
V L Pisacane1, Q E Dolecek, A B Rosenfeld
1vincepisacane@gmail.com
Abstract:
The tissue equivalent proportional counter (TEPC) that utilises a gas cavity has been the standard to obtain microdosimetric observations. An alternative is the solid-state microdosimeter that replaces the gas with a solid-state detector with microscopic sensitive volumes. Here, we describe the development of two versions of a personal solid-state microdosimeter for space exploration applications and give test results for iron and proton beams with comparisons to TEPC measurements and Geant4 radiation transport code simulations. In addition, we describe and provide test results of an optical technique to carry out an end-to-end system test and calibration of a silicon solid-state microdosimeter. This technique eliminates the need for an ionising radiation source with its attendant issues on use and transportation and provides an advantage over the TEPC.
More Related Videos
14:53In Situ Detection and Single Cell Quantification of Metal Oxide Nanoparticles Using Nuclear Microprobe Analysis
Published on: February 3, 2018
06:53Magnetometric Characterization of Intermediates in the Solid-State Electrochemistry of Redox-Active Metal-Organic Frameworks
Published on: June 9, 2023