Specificity characteristics of 7 commercial creatinine measurement procedures by enzymatic and Jaffe method
Neil Greenberg1, William L Roberts, Lorin M Bachmann
1Ortho Clinical Diagnostics, Rochester, NY, USA. ngreenbe@frontier.com
Background:
Standardized calibration does not change a creatinine measurement procedure's susceptibility to potentially interfering substances.
Methods:
We obtained individual residual serum or plasma samples (n = 365) from patients with 19 different disease categories associated with potentially interfering substances and from healthy controls. Additional sera at 0.9 mg/dL (80 μmol/L) and 3.8 mg/dL (336 μmol/L) creatinine were supplemented with acetoacetate, acetone, ascorbate, and pyruvate. We measured samples by 4 enzymatic and 3 Jaffe commercially available procedures and by a liquid chromatography/isotope dilution/mass spectrometry measurement procedure against which biases were determined.
Results:
The number of instances when 3 or more results in a disease category had biases greater than the limits of acceptability was 28 of 57 (49%) for Jaffe and 14 of 76 (18%) for enzymatic procedures. For the aggregate group of 59 diabetes samples with increased β-hydroxybutyrate, glucose, or glycosylated hemoglobin (Hb A(1c)), the enzymatic procedures had 10 biased results of 236 (4.2%) compared with 89 of 177 (50.3%) for the Jaffe procedures, and these interferences were highly procedure dependent. For supplemented sera, interferences were observed in 11 of 24 (46%) of groups for Jaffe and 8 of 32 (25%) of groups for enzymatic procedures and were different at low or high creatinine concentrations.
Conclusions:
There were differences in both magnitude and direction of bias among measurement procedures, whether enzymatic or Jaffe. The influence of interfering substances was less frequent with the enzymatic procedures, but no procedure was unaffected. The details of implementation of a method principle influenced its susceptibility to potential interfering substances.
More Related Videos
09:32Procedure and Key Optimization Strategies for an Automated Capillary Electrophoretic-based Immunoassay Method
Published on: September 10, 2017
07:15Measurement of Cyclic Guanosine Monophosphate (cGMP) in Solid Tissues using Competitive Enzyme-Linked Immunosorbent Assay (ELISA)
Published on: July 3, 2025
Related Concept Videos
Drug Dosing in Renal Diseases: Measurement of Glomerular Filtration Rate
Drug Dosing in Renal Diseases: Measurement of Serum Creatinine Concentration and Clearance
Renal Drug Clearance: Comparison Between Renal Excretion Methods
Renal clearance is often associated with the renal glomerular filtration rate (GFR), which represents the rate at which plasma is filtered through the glomeruli in the kidney. When drug reabsorption is minimal and there is no active secretion, renal clearance is closely related to the...
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or quantified.
Drug Dosing in Renal Diseases: Estimation of Glomerular Filtration Rate Based on Serum Creatinine Concentration
Serum Studies: Renal Function Tests