Is the blood basophil count sufficiently precise, accurate, and specific?: three automated hematology instruments and
Erik K Amundsen1, Carola E Henriksson, Mette R Holthe
1Blood Cell Research Group, Department of Medical Biochemistry, Oslo University Hospital, Ullevaal, Oslo, Norway.
Abstract:
We compared the performance of the basophil count of 3 hematology instruments with a flow cytometric method (FCM) in which CD123 and CD193 were used as basophil markers. By analyzing 112 patient samples, we found the ADVIA 120 (Siemens Healthcare Diagnostics, Deerfield, IL) and CELL-DYN Sapphire (Abbott Diagnostics, Santa Clara, CA) to underestimate the number of basophils by approximately 50% and the Sysmex XE-2100 (Sysmex, Kobe, Japan) and ADVIA to overestimate the basophil count in some samples with pathologic leukocytes. All 3 instruments had large (25%-50%) analytic within-run coefficients of variation. Compared with the FCM, we found a relatively good correlation for the CELL-DYN basophil count (r = 0.81), an intermediate correlation for the Sysmex (r = 0.64), and a poor correlation for the ADVIA (r = 0.24). When excluding the 52 samples flagged for the presence of pathologic leukocytes, these correlations were found to be 0.84, 0.90, and 0.57, respectively. The basophil count of the 3 instruments is, at least presently, of unsatisfactory quality.
More Related Videos
08:17A Semi-automated Approach to Preparing Antibody Cocktails for Immunophenotypic Analysis of Human Peripheral Blood
Published on: February 8, 2016
14:45Enumeration of Major Peripheral Blood Leukocyte Populations for Multicenter Clinical Trials Using a Whole Blood Phenotyping Assay
Published on: September 16, 2012