Magnetically-Assisted Remote Controlled Microcatheter Tip Deflection under Magnetic Resonance Imaging
Published on: April 4, 2013
Catheter tracking with phase information in a magnetic resonance scanner
Kevan J T Anderson1, Greig C Scott, Graham A Wright
1Department of Medical Biophysics, University of Toronto, Toronto, ON M5S 1A1, Canada. kevan.anderson@sri.utoronto.ca
Abstract:
The purpose of this study is to describe a new active technique for accurately determining both the position and orientation of the tip of a catheter during magnetic resonance (MR)-guided percutaneous cardiovascular procedures. The technique utilizes phase information introduced into the MR signal from a small receive coil located on the distal tip of the catheter. Phase patterns around a small receive coil are rich in information that is directly related to position and orientation. This information can be collected over a large spherical volume with a diameter several times that of the receive coil. The high degree of redundancy yields the potential for an accurate and robust method of catheter tracking. A tracking algorithm is presented that performs catheter tip localization using phase images acquired in two orthogonal planes without any a priori knowledge of catheter position. Associated experimentation demonstrating feasibility is also presented.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...