Related Experiment Video
Updated: May 26, 2026

Design of a Cyclic Pressure Bioreactor for the Ex Vivo Study of Aortic Heart Valves
Published on: August 23, 2011
Posture-independent piston valve: a novel valve mechanism that actuates based on intracranial pressure alone
Joshua E Medow1, Christopher C Luzzio
1Department of Neurosurgery, University of Wisconsin School of Medicine and Public Health, Madison, Wisconsin, USA. medow@neurosurg.wisc.edu
Object:
Shunt valves are intended to maintain physiological intracranial pressure (ICP). A variety of mechanisms have been designed to accomplish this goal but have had limited success. Siphoning, in particular, has been a problem not effectively solved by proposed or manufactured valves. Poor control of ICP results in headache, neurological disturbances, decreased cognition, shunt malfunction, slit ventricles, subdural hematomas, decreased cranial volume, and maldevelopment. The authors of this study describe a prototype valve that was machined and tested ex vivo and that actuates based on ICP alone regardless of the presence of a siphon. Their object was to determine if a novel shunt mechanism that actuates perpendicular to the flow of fluid would eliminate the effect of siphoning in a valve for the treatment of hydrocephalus.
Methods:
A posture-independent piston valve (PIPV) was anchored to a graduated reservoir. Opening pressure was measured by noting the fluid level in the reservoir when the piston moved. Measurements were made using a 90-cm and a 120-cm water-filled siphon tube (1.3-mm standard distal catheter) to simulate an upright posture. A recumbent posture was simulated by the absence of a siphon.
Results:
Opening pressure of the valve did not change regardless of the presence or absence of a water-filled siphon.
Conclusions:
The PIPV was triggered only by the pressure head at the inlet and did not actuate in the presence of a siphon, demonstrating proof of principle of the perpendicular actuating mechanism. The PIPV is a purely mechanical device that has practical application in the treatment of hydrocephalus.
Related Concept Videos
Application of Pascal's Law
Hydraulic systems are used to operate automotive brakes, hydraulic jacks, and numerous other mechanical systems. We can derive a relationship between the forces in a simple hydraulic system by applying...
Increased Intracranial Pressure l: Introduction
Increased Intracranial Pressure ll: Pathophysiology
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...