Accuracy and precision of the Prodigy AutoCode blood glucose monitor
Emily S Prohaska1, Charles Herring, Gregory B Russell
1The University of Kansas Hospital, Kansas City, KS 66160, USA. eprohaska@kumc.edu
Objectives:
To assess the accuracy and precision of the Prodigy AutoCode blood glucose monitor.
Methods:
This open-label, prospective equivalence study was conducted at Wilson Community Health Center in Wilson, North Carolina. Accuracy was assessed by comparing finger stick blood glucose values to venipuncture. Precision was assessed by comparing consecutive finger stick blood glucose values from 2 Prodigy AutoCode monitors. Data were analyzed using paired t tests, signed rank tests, regression, and mixed effect models.
Results:
Fifty-three subjects completed the study. Meter 1 produced 14 (26%) and meter 2 produced 13 (25%) blood glucose readings outside the acceptable error margin of ±20% set by the International Organization for Standardization (ISO) and the Food and Drug Administration (FDA). Neither meter was accurate compared to venipuncture (P < .001 for both). Consecutive blood glucose results obtained from meters 1 and 2 were precise (meter 1 vs 2, P = .533).
Conclusions:
The Prodigy AutoCode demonstrated precision between two different monitors but was inaccurate compared to venipuncture. Less than the required 95% of blood glucose values from each monitor fell within the acceptable 20% error margin relative to venipuncture. All readings outside the acceptable error margin were overestimations, indicating potentially significant safety concerns including untreated or undertreated hypoglycemia.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Accuracy and Precision
Accuracy and Precision
Uncertainty in Measurement: Accuracy and Precision
Uncertainty in Measurement: Reading Instruments
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...