Agreement of two different laboratory methods used to measure electrolytes
Venencia Albert1, Arulselvi Subramanian, Kanchana Rangarajan
1Department of Laboratory Medicine, JPNATC, AIIMS, New Delhi, India.
Aim:
The aim of our study was to do an agreement analysis of two different laboratory methods used to measure electrolytes i.e., between the ISE based Beckman Coulter Synchron CX9 PRO Biochemistry analyzer and RAL's Ion3 Flame Photometer (Técnica para el Laboratorio, Barcelona, Spain), in serum samples.
Materials And Methods:
This cross sectional study was done over a period of three months from September'09 through December'09 on routine biochemistry samples. A total of 6492 samples were received for routine biochemistry analysis from those 630 blood samples were randomly processed for this study. Two ml of sample was taken in a plain gel tube (LABTECH Disposables, Ahmedabad, India), centrifuged and further processed using both systems within one hour of the sampling to obtain the Na and K concentrations in the samples. The bias and variability of differences in measured values were analyzed according to Bland and Altman method.
Results:
Flame photometry method has drawbacks such as low throughput, requires manual operation, is a time consuming procedure. Ion selective electrodes technique is a more universal method for the high throughput determination of electrolytes in physiological samples; Beckman Coulter Synchron CX9 PRO is an example of such a system. The mean difference between the two methods (standard minus test) and 95% limits of agreement for sodium in serum was -7.8±17.3 (-42.2 to 26.6) and in urine was -22±41 (-104 to 60). Similarly, the mean difference between the two methods for potassium values in serum was found to be -0.25±0.75 (-1.75 to 1.25) and in urine was -5.3±38.9 (-83.1 to 72.5). With 95% confidence interval, the value of sodium and potassium as determined by both the methods lie between the upper and lower limit showing 95% limits of agreement.
Conclusion:
Good degree of agreement was seen on comparing the two methods for measuring the electrolytes; the use of Synchron CX9 in place of Flame photometer for electrolyte analysis in serum and urine is justified or use the two interchangeably.
Related Concept Videos
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
Potentiometry: Overview
Interfacial Electrochemical Methods: Overview
Electrogravimetric Analysis: Overview
To test the completeness of the...
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Coulometry: Overview