Comparison of four pulse oximeters: effects of venous occlusion and cold-induced peripheral vasoconstriction
J A Langton1, D Lassey, C D Hanning
1University Department of Anaesthesia, Leicester General Hospital.
Abstract:
The ability of four pulse oximeters (the Ohmeda 3700, Nellcor N100 and N200 and the Datex Oscar) to detect hypoxaemia was determined in the presence of venous obstruction and cold-induced peripheral vasoconstriction. Significant increases in detection time for hypoxaemia were found in both cases. There were no significant differences in detection time between the instruments, although the Ohmeda 3700 displayed smaller values of SaO2 under certain conditions. Peripheral vasoconstriction was induced using three differing methods which gave differing results, thus emphasizing the importance of methodology in assessments of pulse oximetry.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
04:44Near-Infrared Spectroscopy During Reactive Hyperemia for the Assessment of Lower Limb Vascular Function
Published on: March 22, 2024
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Measurement of Blood Pressure
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above the...