Employing the Forced Oscillation Technique for the Assessment of Respiratory Mechanics in Adults
Published on: February 9, 2022
Accuracy and precision of desktop spirometers in general practices
Tjard R J Schermer1, Eddy H A Verweij, Riet Cretier
1COPD and Asthma Research and Development Unit, Department of Primary and Community Care, Radboud University Nijmegen Medical Centre, Nijmegen, The Netherlands.
Desktop spirometers in general practices show average accuracy for lung function tests, but some devices have significant errors. Regular calibration is crucial for reliable spirometry results in diagnosing and managing airway diseases.
Area of Science:
- Pulmonary Medicine
- Medical Device Performance
- Diagnostic Accuracy
Background:
- Spirometry is vital for diagnosing and monitoring chronic airway diseases in general practices.
- The performance and accuracy of commonly used spirometry equipment in primary care settings are not well-understood.
- Inaccurate spirometry devices can negatively impact patient diagnosis and treatment.
Purpose of the Study:
- To assess the accuracy and precision of desktop spirometers utilized in routine general practice.
- To identify potential factors influencing spirometer performance, such as device type and age.
Main Methods:
- A random sample of 50 desktop spirometers from Dutch general practices was evaluated.
- Devices were tested using a certified waveform generator with 8 standard American Thoracic Society (ATS) waveforms.
- Data on device type, age, usage, cleaning, and calibration were collected via questionnaire.
Main Results:
- Turbine spirometers comprised 80% of the tested devices; pneumotachographs represented 16%.
- Average deviations from reference values were 25 ml for FEV1 and 27 ml for FVC, with some devices exhibiting substantial inaccuracies.
- Pneumotachographs showed larger FEV1 deviations compared to turbine spirometers; no significant difference was found for FVC.
Conclusions:
- Desktop spirometers in general practices tend to slightly overestimate FEV1 and FVC, with notable deviations in some units.
- The findings underscore the need for general practices to enhance their focus on spirometer calibration.
- Ensuring accurate spirometry is essential for reliable patient care in primary settings.
Related Concept Videos
Pulmonary Function Tests
Pulmonary Function Tests are crucial diagnostic tools for assessing respiratory function, particularly in patients with chronic respiratory disorders. They comprehensively evaluate lung volumes, ventilatory function, breathing mechanics, diffusion, and gas exchange. These tests help diagnose pulmonary diseases and play a significant role in monitoring disease progression, evaluating disability, and assessing response to therapy.
PFTs involve using a spirometer, a...
Uncertainty in Measurement: Accuracy and Precision
Errors occurring during blood pressure monitoring
Several factors...
Chronic Obstructive Pulmonary Disease-IV: Assessement and Diagnostic Studies
Medical History
Assessment of blood pressure in brachial artery(two-step method)
Accuracy and Precision