In-line filter included into the syringe infusion pump assembly reduces flow irregularities
1Department of Neonatology and Intensive Care, University Children's Hospital Zurich, Steinwiesstr 75, 8032 Zurich, Switzerland. barbara.brotschi@kispi.uzh.ch
Purpose:
To evaluate whether an in-line filter inserted in the syringe pump infusion line assembly influences start-up times and flow irregularities during vertical pump displacement at low infusion rates.
Methods:
Fluid delivery after syringe pump start-up and after vertical displacement of the syringe pump by -50 cm was determined gravimetrically at flow rates of 0.5, 1.0 and 2.0 ml h(-1). Measurements were repeated for each flow rate four times with two different syringe pumps with and without an in-line filter incorporated. Data are shown as median and range.
Results:
Start-up times were reduced by an in-line filter at 0.5 ml h(-1) flow rate from 355.5 s (0-660) to 115 s (0-320), whereas the effect was attenuated at higher flow rates. Pooling of fluid into the infusion system after lowering the infusion syringe pump was halved in all flow rates tested. Amount of infusion bolus after elevating the syringe pump by 50 cm was not affected by an in-line filter.
Conclusion:
In the evaluated model in-line filters help to reduce flow irregularities and delay in drug delivery of syringe pumps at low flow rates and represent an option to optimize continuous administration of highly concentrated short-acting drugs at very small infusion rates.
More Related Videos
09:08Three-dimensional Printing of Thermoplastic Materials to Create Automated Syringe Pumps with Feedback Control for Microfluidic Applications
Published on: August 30, 2018
07:30In Vitro Model of Physiological and Pathological Blood Flow with Application to Investigations of Vascular Cell Remodeling
Published on: November 3, 2015
Related Concept Videos
Drug Accumulation During Multiple Dosing: Intermittent IV Infusions
Filtration
Dialysis