Yunlai Wang1, Xiongfei Liao, Ruigang Ge
1Department of Radiation Oncology, PLA General Hospital, 100853 Beijing, China. nanyang@163.com
Objective:
To evaluate the radiation output and stability of linac-integrated kV cone beam CT unit.
Methods:
Air kermas in radiographic mode were measured with 0.6 cc ion chamber and Unidos electrometer for Synergy-integrated XVI kV cone beam CT unit. Air kermas vs image frames were measured in fluoroscopic mode. Output stability and depth doses were measured.
Results:
The air kerma increased quadratically with the increased tube voltage, while increasing linearly with the tube current, exposure time, and number of frames. The radiation output stability and its change with the gantry angle were within +/-1%. The percentage depth dose increased with higher tube voltage.
Conclusion:
The radiation output of XVI is stable. The radiation outputs change considerably with the preset parameters. Parameters should be optimally chosen to reduce the patient dose.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies for Cardiovascular System V: CT
X-ray Imaging