Pre-clinical evaluation of titanium nitride coated titanium material
C S Geetha1, A Sabareeswaran, P V Mohanan
1Biomedical Technology Wing, Sree Chitra Tirunal Institute for Medical Sciences and Technology, Kerala, India.
Toxicology Mechanisms and Methods
|January 17, 2012
Summary
Titanium nitride coated titanium, a material for left ventricular assist devices, underwent rigorous biocompatibility testing. The material proved non-toxic, non-irritant, and non-haemolytic, meeting international standards for medical device fabrication.
Area of Science:
- Biomaterials Science
- Medical Device Engineering
- Toxicology
Background:
- Titanium nitride (TiN) coated titanium is a candidate material for left ventricular assist devices (LVADs).
- Pre-clinical evaluation according to international standards is crucial before fabricating medical devices.
- Assessing the biocompatibility and toxicity of TiN coated titanium is essential for its safe application in LVADs.
Purpose of the Study:
- To evaluate the biocompatibility and toxicity of titanium nitride coated titanium.
- To determine if the material meets international standards for medical device fabrication.
- To assess potential risks including acute systemic toxicity, irritation, hemolysis, and tissue response.
Main Methods:
- Acute systemic toxicity tests using physiological saline and cottonseed oil extracts in mice.
- Intracutaneous irritation tests in rabbits with material extracts.
- In vitro hemolysis assays using rabbit blood.
- Muscle implantation studies in rabbits followed by histopathological analysis at 1, 4, and 12 weeks.
Main Results:
- No significant systemic toxicity or irritation was observed with material extracts.
- Hemolysis percentages were within acceptable ranges, indicating minimal impact on red blood cells.
- Histopathological analysis revealed no adverse cellular changes at the implantation sites.
Conclusions:
- Titanium nitride coated titanium demonstrates favorable biocompatibility.
- The material is non-toxic, non-irritant, and non-haemolytic.
- The findings support the use of TiN coated titanium for left ventricular assist device fabrication.