An In Vitro Hemodynamic Loop Model to Investigate the Hemocytocompatibility and Host Cell Activation of Vascular Medical Devices
Published on: August 21, 2020
Replication of the Frank-Starling response in a mock circulation loop
Shaun D Gregory1, Michael Stevens, Daniel Timms
1Faculty of Built Environment and Engineering and Institute of Health and Biomedical Innovation, Queensland University of Technology, Brisbane, Australia. shaun.gregory@qut.edu.au
This study introduces a simulated Frank-Starling response into mock circulation loops (MCLs) to better mimic human cardiovascular autoregulation. This enhancement improves device testing and reduces the need for animal trials.
Area of Science:
- Cardiovascular physiology
- Biomedical engineering
- Medical device testing
Background:
- Mock circulation loops (MCLs) are essential for pre-clinical cardiovascular device evaluation.
- Current MCLs lack human autoregulatory responses, limiting their predictive accuracy.
- Accurate simulation of cardiac function is crucial for reliable device testing.
Purpose of the Study:
- To develop and implement a functional left and right ventricular Frank-Starling response within an MCL.
- To enhance MCLs with human-like autoregulation for more realistic cardiovascular simulations.
- To validate the simulated Frank-Starling response against established human physiological data.
Main Methods:
- A proportional controller regulated pneumatic ventricular driving pressure based on end-diastolic volume.
- Ventricular pressure-volume loops were generated under various physiological conditions.
- Simulated cardiac contractility was assessed by analyzing end-systolic pressure-volume relationships.
Main Results:
- A non-linear Frank-Starling response was successfully implemented in the MCL.
- The simulated response demonstrated accurate alterations in ventricular contractility with changing preload.
- Generated data showed strong agreement with previously reported human physiological data.
Conclusions:
- The enhanced MCL accurately simulates the Frank-Starling mechanism, a key aspect of human cardiac function.
- This improved test rig offers a more physiologically relevant platform for cardiovascular device refinement.
- The study successfully reduced the reliance on in-vivo trials by providing a validated simulation.
More Related Videos
07:30In Vitro Model of Physiological and Pathological Blood Flow with Application to Investigations of Vascular Cell Remodeling
Published on: November 3, 2015
08:50Gene Expression Analysis of Endothelial Cells Exposed to Shear Stress Using Multiple Parallel-plate Flow Chambers
Published on: October 21, 2018
Related Concept Videos
Autoregulation of Blood Flow
Chemical Signaling in Autoregulation
Chemical signaling operates at the precapillary sphincter level, inciting either contraction or relaxation.
Regulation of Stroke Volume
Preload refers to the degree of stretch on the heart before it contracts. It's analogous to the stretching of a rubber band; the more it's stretched, the more forcefully it snaps back. This concept is encapsulated in the Frank-Starling law of the...