Effects of hydrofluoric acid and anodised micro and micro/nano surface implants on early osseointegration in rats
Yongfeng Li1, Yuan Gao, Bo Shao
1Department of Oral and Maxillofacial Surgery, School of Stomatology, the Fourth Military Medical University, 145 West Changle Road, Xi'an 710032, Shaanxi, China.
Abstract:
Our aim was to evaluate the effects of hydrofluoric acid and anodised micro and micro/nano surface implants on bony ingrowth in the earliest stage of implantation in rats. Sixty cylindrical screwed titanium alloy implants with machined, micro, and hierarchical hybrid micro/nano surfaces (n=20 in each group) were inserted into the distal femurs of 30 female Sprague-Dawley rats. In vivo microcomputed tomography (micro CT) was used to assess microarchitectural changes in the bone around the implants 2 weeks after implantation. All the animals were then killed and the femurs with implants harvested for histological analysis and pull-out testing. Micro CT analysis showed that the trabecular thickness and the bone:volume ratio (bone volume:total volume) (BV:TV) increased significantly in the micro/nano group compared with the other two groups, while the trabecular separation decreased significantly in the micro/nano group compared with the machined group. The mean (SD) bone-implant contacts (%) were 38.94 (9.48), 41.67 (8.71), and 51.49 (12.49) in the machined, micro, and micro/nano groups, respectively. The maximum pull-out forces (N) were 64.95 (6.11), 71.45 (7.15), and 81.90 (13.1), respectively. Both bone-implant contacts and maximum pull-out forces were significantly higher in the micro/nano group, but there was no significant difference between the micro group and the machined group. These data indicate that the hierarchical hybrid micro/nano surface of the implant can promote osseointegration in the earliest stage of implantation, and may be a promising option for further clinical use.