Related Experiment Video
Updated: May 25, 2026

Conducting Respiratory Oscillometry in an Outpatient Setting
Published on: April 8, 2022
Accuracy of office spirometry performed by trained primary-care physicians using the MIR Spirobank hand-held
J Degryse1, J Buffels, Y Van Dijck
1Department of Primary Health Care, KU Leuven, Leuven, Belgium. Jan.degryse@med.kuleuven.be
Background:
With the availability of compact, portable, effective microspirometers, pulmonary function tests no longer need to be performed only in specialized laboratories. However, the perception persists that small flow-sensing devices are less accurate than volume-sensing spirometers.
Objectives:
To study the accuracy of spirometry performed with the MIR Spirobank® and to investigate how accurately trained primary-care physicians can perform spirometry using a portable electronic spirometer.
Methods:
Patients with suspected occupational asthma were submitted to specific bronchial challenge tests in the pulmonary function laboratory according to published recommendations. Serial measurements were performed with the Jaeger MasterScope device (reference standard) or the Spirobank device. Data were generated from 908 parallel measurements on 34 patients. Furthermore, 16 patients with documented moderate to severe COPD were examined in a carousel set-up by four trained physicians who each used his/her own Spirobank device coupled to a laptop computer.
Results:
The Spirobank spirometer performed very well compared with the Jaeger MasterScope in a laboratory environment, displaying an underestimation of the forced expiratory volume in 1 s (FEV(1)) and FEV(1)/forced vital capacity (FVC) of 2-5%. High correlations were found for the pulmonary function parameters. The highest correlation was for FEV(1) (r(2) = 0.949) and the lowest for the maximum expiratory flow at 25% of FVC (MEF(25)) (r(2) = 0.864). Only 2% of the observed variation in the measurement results could be explained by the type of device.
Conclusions:
The Spirobank device seems to be appropriate for research purposes if the standardized protocol is used correctly and the acceptability criteria are respected.
Related Concept Videos
Pulmonary Function Tests
Pulmonary Function Tests are crucial diagnostic tools for assessing respiratory function, particularly in patients with chronic respiratory disorders. They comprehensively evaluate lung volumes, ventilatory function, breathing mechanics, diffusion, and gas exchange. These tests help diagnose pulmonary diseases and play a significant role in monitoring disease progression, evaluating disability, and assessing response to therapy.
PFTs involve using a spirometer, a...
Chronic Obstructive Pulmonary Disease-IV: Assessement and Diagnostic Studies
Medical History
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Errors occurring during blood pressure monitoring
Several factors...
Assessment of Respiration
Subjective Assessment: Nurses interview the patient to gather information directly during the subjective assessment. It includes questions about the individual's medical history, medications, and symptoms, focusing on past respiratory conditions like asthma or COPD,...