Related Experiment Video
Updated: May 25, 2026

Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Calculation of beam quality correction factors for various thimble ionization chambers using the Monte Carlo code
Fabián Erazo1, Antonio M Lallena
1Departamento de Física Atómica, Molecular y Nuclear, Universidad de Granada, E-18071 Granada, Spain.
Abstract:
The beam quality correction factor [Formula: see text] and the perturbation factor pQ, commonly considered in dosimetry with ionization chambers, were calculated for the NE2571 and the Standard Imaging A19 and A12S chambers, using the Monte Carlo simulation code PENELOPE. For the NE2571 chamber, the values of [Formula: see text] obtained are in very good agreement with those found in previous works by Wulff et al. and Muir and Rogers with the code EGSnrc and also with the experimental results summarized in the NCS code of practice. For pQ, a difference of ∼0.4% has been found between our results and those obtained with EGSnrc for (60)Co and this difference increases slightly with TPR10(20) values. These factors have been calculated also for the A19 and A12S chambers of Standard Imaging. The values of [Formula: see text] show reasonable agreement with those recently calculated by Muir and Rogers and the measurements of McEwen.
