A first evaluation of breast radiological density assessment by QUANTRA software as compared to visual classification
Stefano Ciatto1, Daniela Bernardi, Massimo Calabrese
1UO Senologia Clinica e Screening Mammografico, Dipartimento di Radiodiagnostica, APSS, Trento, Italy. stefano.ciatto@gmail.com
Computerized breast density assessment offers superior reproducibility compared to visual methods. This automated approach aids in standardizing mammography interpretation for personalized breast cancer screening.
Area of Science:
- Radiology
- Medical Imaging
- Oncology
Background:
- Breast density is a key factor in breast cancer risk and mammogram effectiveness.
- Personalized screening strategies can be informed by breast density assessments.
- Variability in visual density assessment poses challenges for consistent clinical application.
Purpose of the Study:
- To evaluate the reproducibility of visual breast density assessment by radiologists.
- To compare computer-assisted breast density measurements with visual assessments.
- To establish a reliable method for computer-assessed breast density in clinical practice.
Main Methods:
- Eleven experienced radiologists visually classified 418 digital mammograms using BI-RADS categories.
- The QUANTRA software was used for computer-assisted breast density assessment.
- Statistical analysis, including kappa values and cut-off analysis, was performed to compare methods.
Main Results:
- Visual density assessment showed satisfactory reproducibility (kappa 0.620-0.851) but significant inter-reader variability (25.1-50.5% dense breasts).
- QUANTRA software yielded systematically lower density percentages than visual assessment.
- A QUANTRA cut-off of ≤22.0% accurately predicted visual classification into dense and non-dense categories (89.0% overall accuracy).
Conclusions:
- Computer-assessed breast density is highly reproducible and preferable to subjective visual assessment.
- The QUANTRA software, with a threshold of ≤22.0%, can reliably differentiate dense from non-dense breasts.
- Further comparative studies are needed to refine computer-assessed density for clinical use.