MR safety: fast T₁ thermometry of the RF-induced heating of medical devices
1Research Center for Magnetic Resonance Bavaria e.V., Würzburg, Germany. daniel.gensler@physik.uni-wuerzburg.de
Abstract:
Determining the MR compatibility of medical implants and devices is becoming increasingly relevant. In most cases, the heating of conductive implants due to radiefrequency (RF) excitation pulses is measured by fluoroptic temperature sensors in relevant tests for approval. Another common method to determine these heating effects is MR thermometry using the proton resonance frequency. This method gives good results in homogeneous phantoms. However in many cases, technical shortcomings such as susceptibility artifacts prohibit exact proton resonance frequency thermometry near medical implants. Therefore, this work aimed at developing a fast T₁-based method which allows controlled MR-related heating of a medical implant while simultaneously quantifying the spatial and temporal temperature distribution. To this end, an inversion recovery snapshot Fast Low-Angle Shot (FLASH) sequence was modified with additional off-resonant heating pulses. With an accelerated imaging method and a sliding-window technique, every 7.6 s a new temperature map could be generated with a spatial in-plane resolution of 2 mm. The temperature deviation from calculated temperature values to reference fluoroptic probe was found to be smaller than 1 K.
More Related Videos
09:01Magnetic-, Acoustic-, and Optical-Triple-Responsive Microbubbles for Magnetic Hyperthermia and Pothotothermal Combination Cancer Therapy
Published on: May 22, 2020
08:52Near-Infrared Temperature Measurement Technique for Water Surrounding an Induction-heated Small Magnetic Sphere
Published on: April 30, 2018
Related Concept Videos
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Absorption of Radiation
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...