Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
Metal artifact reduction for CT: development, implementation, and clinical comparison of a generic and a
Raoul M S Joemai1, Paul W de Bruin, Wouter J H Veldkamp
1Department of Radiology, Leiden University Medical Center, Leiden, The Netherlands. r.m.s.joemai@lumc.nl
Purpose:
To develop, implement, and compare two metal artifact reduction methods for CT.
Methods:
Two methods for metal artifact reduction were developed. The first is based on applying corrections in a Radon transformation of the CT images. The second method is based on a forward projection of the CT images and applying corrections in the scanner's original raw data. The first method is generic since it does not depend on the scanner specifications. For the second method, detailed information on the design of the CT scanner and the raw data of the study is required. Clinical implementation and evaluation were performed using pre- and post-operative CT scans of four patients with shoulder prosthesis. For comparison of these methods, the authors developed a quantitative technique that compares improvement in image quality for the two metal artifact reduction techniques with the image quality of the uncorrected images.
Results:
Metal artifact reduction using either of the two methods yields a decrease of noise and artifacts in CT scans of patients with shoulder prostheses. Artifacts that appeared as bright and dark streaks were reduced or eliminated and as a result image quality improved. Quantitative assessment of clinical images showed improved image quality for both techniques of metal artifact reduction, but the method based on correction in original raw data performed better in all comparisons.
Conclusion:
Both methods are effective for metal artifact reduction, but better performance was observed for the method that is based on correcting the original raw data. The used evaluation technique provides an objective way of evaluating the metal artifacts in clinical CT images.
More Related Videos
08:41Novel In Vivo Micro-Computed Tomography Imaging Techniques for Assessing the Progression of Non-Alcoholic Fatty Liver Disease
Published on: March 24, 2023
10:24Neutron Radiography and Computed Tomography of Biological Systems at the Oak Ridge National Laboratory's High Flux Isotope Reactor
Published on: May 7, 2021
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Magnetic Resonance Imaging
Imaging Studies III: Computed Tomography