Pacemaker/implantable cardioverter-defibrillator dose in balloon high-dose-rate brachytherapy for breast cancer treatment
Related Experiment Video
Updated: May 25, 2026

Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
Yongbok Kim1, Youssef Arshoun, Mark G Trombetta
1Department of Radiation Oncology, Allegheny General Hospital, Pittsburgh, PA 15212-4772, USA. ykim@wpahs.org
Purpose:
To retrospectively report pacemaker (PM)/implantable cardioverter-defibrillator (ICD) dose in balloon high-dose-rate (HDR) brachytherapy and provide distance-dose graph and table to approximately estimate the maximal device dose.
Methods And Materials:
For 3 patients (A, B, and C), PM/ICD was retrospectively contoured on planning computed tomography images and its maximal dose was extracted from a dose-volume histogram. The surface of 1cm expansion from balloon was prescribed to 34Gy and the inverse square law was dominant factor in dose calculation. Therefore, the maximal PM/ICD dose was approximately estimated from the distance-dose graph or table and compared with that of the treatment plan.
Results:
The minimal device-balloon surface distance was 10.9, 18.4, and 4.3cm for patient A, B, and C, respectively. The maximal dose estimated from the proposed table/graph was 2.1 vs. 1.61Gy for patient A, 0.87 vs. 0.49Gy for patient B, and 8.9 vs. 9.14Gy for patient C compared with that from the treatment plan.
Conclusions:
Depending on the location of PM/ICD relative to the tumor bed, balloon HDR brachytherapy is feasible if the maximal dose is less than or equal to the dose limit. The proposed distance-dose graph and/or table enable to approximately predict the maximal device dose based on the measurement of minimal distance between lumpectomy and the device before balloon implantation for the suitability of balloon HDR brachytherapy.