Study on the microbial safety of an infusion set for contrast-enhanced imaging
Marlein Miranda Cona1, Matthias Bauwens, Yichao Zheng
1Radiology Section, Department of Imaging and Pathology, Biomedical Sciences Group, University of Leuven, Belgium.
Objectives:
Multiple uses of automatic contrast injection systems may impose septic risks on patients. The purpose of this experiment was to verify whether a newly developed replaceable patient-delivery system may allow multiple uses of the system but without such risks.
Methods:
Twelve patient-delivery systems were tested according to a multiple-use approach using an automatic contrast injection system consisting of dual syringes and one filling and injecting set. Two protocols with normal saline only (n = 6) or contrast media plus normal saline (n = 6) loaded in the injection system were performed. Each patient-delivery system was connected through an infusion catheter to the ear vein of a rabbit that was intravenously preinjected with a diffusible radiotracer (99m)Tc-dimercaptopropionyl-human serum albumin. Aliquots were sampled from the filling and injecting set, patient line, and animal blood for radioactive analysis after the replacement of each patient-delivery system.
Results:
For the protocol performed using only normal saline, radioactivity was found in the blood circulation of the rabbit (1655903 ± 593221 CPM) and in the patient line (52894 ± 33080 CPM), but, virtually, in none of samples from the filling and injecting set (8 ± 3 CPM), relative to the background (7 ± 3 CPM) (P = 0.726). Similarly, experimental results attained using contrast plus saline show radioactivity in the blood circulation of the rabbit (1119107 ± 183174 CPM) as well as in the patient line (32991 ± 20232 CPM) but in none of samples from the filling and injecting set (6 ± 6 CPM), relative to the background (6 ± 4 CPM) (P = 0.955).
Conclusions:
The tested patient-delivery system proves convenient and safe. It allows multiple uses of the contrast injection system and avoids the risk of cross contamination.
More Related Videos
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies VII: Vascular Imaging
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System IV: CMRI