Using a monochromator to improve the resolution in TEM to below 0.5Å. Part I: Creating highly coherent monochromated
P C Tiemeijer1, M Bischoff, B Freitag
1FEI Company, P.O. Box 80066, 5600 KA Eindhoven, The Netherlands. p.tiemeijer@fei.com
Abstract:
Chromatic aberration limits the resolution in spherical-aberration corrected Transmission Electron Microscopy to approximately 0.7Å at 300 kV. The energy spread in the beam is the main contribution to the chromatic aberration. This spread can be reduced with a monochromator. Another limitation to the resolution in TEM can be the finite brightness of the source and the consequent partial spatial coherence of the illumination. This limitation becomes important when spherical aberration and/or defocus are present such as in uncorrected TEM or in focal-series reconstruction in TEM. We used a monochromator optimized for minimum brightness loss and a prototype 'high-brightness' gun, and obtained brightness after monochromation comparable to that of the standard Schottky FEG before monochromation. The images were acquired on the prototype TEAM 0.5 microscope, which was developed on a Titan platform by increasing its electrical and mechanical stability.
Related Concept Videos
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Super-resolution Fluorescence Microscopy
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Overview of Electron Microscopy
Transmission Electron Microscopy
Confocal Fluorescence Microscopy