The ocular pulse amplitude as a noninvasive parameter for carotid artery stenosis screening: a test accuracy study
Pascal B Knecht1, Moreno Menghini, Lucas M Bachmann
1University Hospital Zurich, Department of Ophthalmology, Zurich, Switzerland. pascal.knecht@usz.ch
Purpose:
To investigate a potential correlation between the ocular pulse amplitude (OPA; i.e., the intraocular pressure difference between the systolic and diastolic phases of the heartbeat) and the severity of carotid artery stenosis (CAS) and to test its role as a screening parameter for CAS during routine ophthalmic examination.
Design:
Test accuracy study.
Participants:
Patients referred for color duplex ultrasound examination of the extra- and intracranial cerebral arteries were enrolled consecutively.
Methods:
We measured OPA on both eyes by dynamic contour tonometry. Multivariate analyses were performed with risk factors for CAS (age, total cholesterol, low-density lipoprotein, and triglycerides) to compare the diagnostic value of OPA measurements with other non- or minimally invasive screening parameters.
Main Outcome Measures:
The difference between OPA measurements in patients with no (<50%) and patients with severe CAS (>70%) as well as the value of OPA measurements to predict the severity of CAS taking further risk factors of CAS into consideration.
Results:
One hundred thirty-four eyes of 67 patients (25 women, 42 men) with a mean age of 67±13 years (range, 25-87) were included. The means of the OPA values of those patients showing no CAS (<50%) differed significantly (P = 0.036) from those with a stenosis of ≥70%. The multivariate model produced a statistically significant odds ratio (0.46; P = 0.007) for CAS of ≥70%.
Conclusions:
The results of the present study provide proof of principle that the OPA is reduced in patients with CAS and may be used as a noninvasive, inexpensive, readily available, and unconfounded screening parameter to detect CAS and possibly to reduce the incidence of stroke.
Financial Disclosure(S):
The authors have no proprietary or commercial interest in any of the materials discussed in this article.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Pulse
The pulse serves as a clinical indicator...
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Pulse amplitude and quality
A weak or absent pulse may indicate reduced cardiac output or poor left ventricular contraction, which can be signs of cardiovascular dysfunction or...