Modification of a draw-over vaporizer for use with sevoflurane
T Payne1, R Neighbour, R Eltringham
1Department of Anaesthesia, Gloucestershire Royal Hospital, Gt Western Rd, Gloucester GL1 3NN, UK. tompayne@doctors.net.uk
Background:
Draw-over anaesthesia is widely used throughout the developing world, in disaster areas and in military anaesthesia when the supply of pressurized oxygen is unreliable. To date, no draw-over vaporizer has been able to deliver sufficient concentrations of sevoflurane for use in inhalation induction of anaesthesia. A laboratory study to assess the performance of a new vaporizer (DDV2) to deliver sevoflurane in a wide range of situations is described.
Methods:
In this study, the concentration of sevoflurane delivered at different dial settings (1-4%) and at different temperatures (20-40°C) in a draw-over mode was measured. The concentration of sevoflurane delivered at different dial settings with continuous flow (6 and 8 litre min(-1)) at 20°C was measured. The maximum possible concentration of sevoflurane that can be delivered by the DDV2 was measured at a continuous flow rate of 8 litre min(-1) at 20, 30, and 40°C.
Results:
Concentrations of sevoflurane delivered in the draw-over mode were within 0.5% of dialled setting up to 30°C. Above this temperature, higher levels of vapour were delivered. With continuous flow, concentrations of sevoflurane at 20°C were within 0.5% of dialled setting and were stable throughout the duration of the experiment. On the 'induction' setting, concentrations of sevoflurane of between 6.4% and 10.1% could be delivered with continuous flow.
Conclusions:
The modifications to the DDV2 allow stable concentrations of sevoflurane to be delivered in draw-over and continuous flow modes over a range of temperatures. With continuous flow, concentrations of sevoflurane sufficient for induction of anaesthesia can be achieved.
More Related Videos
Related Concept Videos
Vaporization
Inhalational Anesthetics: Overview
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Vapor Pressure Lowering
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...