A simple approach for EPID dosimetric calibration to overcome the effect of image-lag and ghosting
Mukhtar Alshanqity1, Simon Duane, Andrew Nisbet
1Department of Physics, CNRP, University of Surrey, and Royal Surrey County Hospital, NHS Foundation Trust, Guildford, UK. mukhtar@physicist.net
Abstract:
EPID dosimetry has known drawbacks. The main issue is that a measurable residual signal is observed after the end of irradiation for prolonged periods of time, thus making measurement difficult. We present a detailed analysis of EPID response and suggest a simple, yet accurate approach for calibration that avoids the complexity of incorporating ghosting and image-lag using the maximum integrated signal instead of the total integrated signal. This approach is linear with dose and independent of dose rate.