Performance validation of a modified magnetic resonance imaging-compatible temperature probe in children
Viviane G Nasr1, Roman Schumann, Iwona Bonney
1Department of Anesthesiology, Tufts Medical Center, Boston, MA 02111, USA. vnasr@tuftsmedicalcenter.org
Introduction:
During magnetic resonance imaging (MRI), children are at risk for body temperature variations. The cold MRI environment that preserves the MRI magnet can cause serious hypothermia. On the other hand, hyperthermia may also develop because of radiofrequency-induced heating of the tissues, particularly in prolonged examinations. Because of a lack of MRI-compatible core temperature probes, temperature assessment is unreliable, and specific absorption rate-related patient heat gain must be calculated to determine the allowable scan duration. We compared an MRI-compatible temperature probe and a modification thereof to a standard esophageal core body temperature probe in children.
Methods:
Children undergoing general anesthesia were recruited, each patient serving as his/her own control. Core body temperature was measured using 3 different devices: (1) a fiberoptic MRI-compatible skin surface temperature probe (MRI-skin) located on the child's skin surface; (2) a fiberoptic MRI-compatible temperature probe modified with a single-use sleeve at the tip (MRI-core), located in the nasopharynx; and (3) a standard temperature monitor (STRD) located in the esophagus or nasopharynx. The Bland-Altman method was used for statistical analysis.
Results:
We enrolled 60 children aged 7.8 ± 6 years (mean ± SD) weighing 32.4 (±26.4) kg. The estimated difference between the STRD and MRI-core measurements of core temperature was 0.06°C (confidence interval [CI]: -0.02, 0.15), and between the STRD and the MRI-skin 1.19°C (CI: 0.97, 1.41). According to the Bland-Altman analysis, the 95% limits of agreement ranged from -0.9 to 3.4 and from -1.3 to 1.2 between the STRD and the MRI-skin probe and the MRI-core probe, respectively.
Discussion:
Our results show good agreement between standard esophageal measurements of core temperature and core temperature measured using a modified MRI-core probe during general anesthesia in a general surgical pediatric population. The ability to accurately assess core temperature in the MRI suite may safely allow longer scan times and therefore reduce repeat anesthetic exposure, improve patient safety, and enhance the quality of care in children.
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...