Related Experiment Video
Updated: May 24, 2026

10:42
Preparing an Isotopically Pure 229Th Ion Beam for Studies of 229mTh
Published on: May 3, 2019
Test bench to commission a third ion source beam line and a newly designed extraction system
T Winkelmann1, R Cee, T Haberer
1Heidelberger Ionenstrahl-Therapie Centrum (HIT), D-69120 Heidelberg, Germany. tim.winkelmann@med.uni-heidelberg.de
The Review of Scientific Instruments
|March 3, 2012
Summary
Heidelberg Ion Beam Therapy Center is integrating a new Supernanogan ion source to produce helium beams for patient treatment. This enhances beam quality for helium, proton, and carbon ions, improving cancer therapy options.
Area of Science:
- Medical Physics
- Particle Accelerator Technology
- Radiation Oncology
Background:
- The Heidelberg Ion Beam Therapy Center (HIT) utilizes electron cyclotron resonance (ECR) ion sources for proton and carbon ion therapy.
- Routine operations since 2006 involve two 14.5 GHz ECR ion sources.
- A need for helium beams for patient treatment necessitates the integration of a new ion source.
Purpose of the Study:
- To commission and validate a new Supernanogan ECR ion source and its associated extraction system and spectrometer line.
- To evaluate the beam quality improvements for helium, proton, and carbon beams using different extraction system configurations.
- To assess the feasibility of integrating the new ion source into the existing HIT production facility.
Main Methods:
- Installation of a third ECR ion source (Supernanogan) on a dedicated test bench at HIT.
- Systematic measurements with varied extraction system setups.
- Characterization of beam quality parameters for helium, proton, and carbon beams.
Main Results:
- Demonstrated successful commissioning of the Supernanogan ion source and its associated systems.
- Quantified improvements in beam quality for helium, proton, and carbon beams through optimized extraction configurations.
- Validation of the new ion source's performance for potential clinical application.
Conclusions:
- The new Supernanogan ion source and extraction system show promise for enhancing beam quality at HIT.
- Successful validation paves the way for integrating helium beam capabilities into clinical practice.
- Further development will focus on seamless integration into the production facility for expanded patient treatment options.