Calibration of KRISS reference ionization chamber for key comparison of (99m)Tc measurement
Jong-Man Lee1, K B Lee, S H Lee
1Korea Research Institute of Standards and Science (KRISS), Republic of Korea. euisodong@gmail.com
Abstract:
KRISS, as the national metrology institute of Korea, has used a reference ionization chamber system to certify the activity of (99m)Tc aqueous sources, but could only recently participate in a comparison exercise by the BIPM (BIPM.RI(II)-K4.Tc-99m) to secure the international equivalence of (99m)Tc radioactivity measurement by way of the BIPM transfer instrument (SIRTI). The KRISS ionization chamber system was calibrated about 100 days before the comparison with a (99m)Tc solution source standardized by the 4πβ(LS)-γ(NaI(Tl)) coincidence counting method. During the comparison, beginning with a higher activity mother solution, the KRISS ionization chamber measured its specific activity without a dilution. The activity of a diluted-solution source was measured by the SIRTI at the same time.