Dry Film Photoresist-based Electrochemical Microfluidic Biosensor Platform: Device Fabrication, On-chip Assay Preparation, and System Operation
Published on: September 19, 2017
Dual photonic-electrochemical lab on a chip for online simultaneous absorbance and amperometric measurements
Olga Ordeig1, Pedro Ortiz, Xavier Muñoz-Berbel
1Instituto de Microelectrónica de Barcelona, IMB-CNM (CSIC), Campus UAB, Bellaterra, Barcelona, Spain.
Abstract:
A dual lab on a chip (DLOC) approach that enables simultaneous optical and electrochemical detection working in a continuous flow regime is presented. Both detection modes are integrated for the first time into a single detection volume and operate simultaneously with no evidence of cross-talk. The electrochemical cell was characterized amperometrically by measuring the current in ferrocyanide solutions at +0.4 V vs gold pseudoreference electrode, at a flow rate of 200 μL min(-1). The experimental results for ferrocyanide concentrations ranging from 0.005 to 2 mM were in good agreement with the values predicted by the Levich equation for a microelectrode inside a rectangular channel, with a sensitivity of 2.059 ± 0.004 μA mM(-1) and a limit of detection (LoD) of (2.303 ± 0.004) × 10(-3) mM. Besides, optical detection was evaluated by measuring the absorbance of ferricyanide solutions at 420 nm. The results obtained therein coincide with those predicted by the Beer-Lambert law for a range of ferricyanide concentrations from 0.005 to 0.3 mM and showed an estimated LoD of (0.553 ± 0.001) × 10(-3) mM. The DLOC was finally applied to the analysis of L-lactate via a bienzymatic reaction involving lactate oxidase (LOX) and horseradish peroxidase (HRP). Here, the consumption of the reagent of the reaction (ferrocyanide) was continuously monitored by amperometry whereas the product of the reaction (ferricyanide) was recorded by absorbance. The DLOC presented good performance in terms of sensitivity and limit of detection, comparable to other fluidic systems found in the literature. Additionally, the ability to simultaneously quantify enzymatic reagent consumption and product generation confers the DLOC a self-verifying capability which in turn enhances its robustness and reliability.
Related Concept Videos
Amperometry: Overview
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.