Related Experiment Video
Updated: May 23, 2026

Purification and Analytics of a Monoclonal Antibody from Chinese Hamster Ovary Cells Using an Automated Microbioreactor System
Published on: May 1, 2019
Quantitative impurity analysis of monoclonal antibody size heterogeneity by CE-LIF: example of development and
David A Michels1, Monica Parker, Oscar Salas-Solano
1Department of Protein Analytical Chemistry, Genentech, South San Francisco, CA 94080-4990, USA. michels.david@gene.com
Abstract:
This paper describes the framework of quality by design applied to the development, optimization and validation of a sensitive capillary electrophoresis-sodium dodecyl sulfate (CE-SDS) assay for monitoring impurities that potentially impact drug efficacy or patient safety produced in the manufacture of therapeutic MAb products. Drug substance or drug product samples are derivatized with fluorogenic 3-(2-furoyl)quinoline-2-carboxaldehyde and nucleophilic cyanide before separation by CE-SDS coupled to LIF detection. Three design-of-experiments enabled critical labeling parameters to meet method requirements for detecting minor impurities while building precision and robustness into the assay during development. The screening design predicted optimal conditions to control labeling artifacts while two full factorial designs demonstrated method robustness through control of temperature and cyanide parameters within the normal operating range. Subsequent validation according to the guidelines of the International Committee of Harmonization showed the CE-SDS/LIF assay was specific, accurate, and precise (RSD ≤ 0.8%) for relative peak distribution and linear (R > 0.997) between the range of 0.5-1.5 mg/mL with LOD and LOQ of 10 ng/mL and 35 ng/mL, respectively. Validation confirmed the system suitability criteria used as a level of control to ensure reliable method performance.

