Purpose:
To quantify the impact of relocation to a purpose built pediatric cardiac catheterization laboratory, on patient radiation dose and fluoroscopy time. To provide guide values for radiation exposure during common structural interventions.
Design:
A retrospective review of common structural cardiac interventions performed over 10-years. The era comprised two 5-year periods before and after relocation using different catheter laboratories. Multivariable analysis adjusted for the following variables: era (pre- and post-move), year, operator seniority, patient age, procedure type.
Setting:
A quaternary referral congenital cardiac centre.
Patients:
All patients <18 yrs (n = 756) in whom one of 6 common structural interventions were undertaken between 2000 and 2009.
Main Outcome Measures:
Radiation dose and fluoroscopy time.
Results:
The move to the new laboratory (latter era) was associated with a dramatic reduction in multivariable-adjusted radiation dose, ranging from 64% (aortic coarctation stenting) to 87% (patent arterial duct closure). There was also a year upon year increase in radiation dose of 5.2% [95% confidence interval (CI): 0.6-10.0%], which persisted after the relocation. However, this was associated with a 5.1% yearly decrease in fluoroscopy time (95% CI: -7.9 to -2.2%).
Conclusion:
Use of state-of-the-art catheter equipment is associated with a dramatic reduction in radiation exposure. However, the surprising finding of year upon year increase in exposure (despite the new equipment) combined with decreased fluoroscopy time suggests a temporal decrease in hardware efficiency. This has major implications for hardware replacement.
More Related Videos
Related Concept Videos
Cardiac Catheterization I: Pre-Procedure Overview
Cardiac Catheterization IV: Nursing Management
Cardiac Catheterization III: Left Heart Catheterization
Cardiac Catheterization II: Right Heart Catheterization
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT